CA10S Handheld Veterinary Sidestream Capnograph VET CO2 Patient Monitor ETCO2

$290.00
In stock
SKU
ff761278-4d0c-4d84-a872-7e1352c45b2c
No offers available.

Introduction

The device adopts sidestream method to measure the animal’s parameter related to EtCO2 and AwRR. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow and indirectly reflect the alveolar ventilation. It can also be paired with blood oxygen probe to detect animal blood oxygen and pulse.
It is applicable for use in pre-hospital emergency , emergency department, ICU, operating room of pet hospitals at all levels.


Features

1)End-tidal CO2 measurement function;
2)Respiration rate measurement function;
3)SpO2 measurement function;
4)Pulse rate measurement function;
5)Optional three units;
6)With audible and visual alarm and silencing functions, the alarm limit can be set;
7)Over-limit alarm function for EtCO2, Respiration rate;
8)Low battery alarm function;
9)With the prompt function of “Checking Filterline”;
10)User can calibrate zero by themselves.
11)With atmospheric pressure and temperature compensation functions
12)Rechargeable lithium battery, up to 8-hour user time;


Performance

End-tidal CO2 measurement:
1)Measurement range: 0 ~150mmHg
2)Resolution: 1mmHg
3)Accuracy: 0~40 mmHg:±2 mmHg
41~70 mmHg:±5% of readings
71~100 mmHg:±8% of readings
101~150 mmHg:±10% of readings
Respiration rate measurement:
1)Measurement range: 2rpm~120rpm
2)Resolution: 1rpm
3)Accuracy: ±1rpm
SpO2 measurement:
1)Measurement range: 30%~100%
2)Resolution: 1%
3)Accuracy: 70 % ~ 100 %:±2 %;
30 % ~ 69 %:undefinition.
Pulse rate measurement:
1)Measurement range: 25 bpm ~ 250 bpm;
2)Resolution: 1bpm;
3)Accuracy: ±2 bpm or ±2%,whichever is greater.
Safety characteristic:
1)Device type: internally powered equipment
2)Degree of protection against electric shock: type BF applied part
3)Degree of protection against ingress of liquid: IPX1
Other parameters:
1)Preheat time: 2min;
2)Response time: < 3s;
3)Alarm sound pressure level: 50 dB;
4)Battery life time: 8hours;
5)Gas sampling rate: 60 ml/min ~ 100 ml/min.


Accessories

User Manual
Sampling tube(Cannula tube with filter cotton)
USB cable


Physical Characteristic

Dimension: 159*72*31 mm
Weight: 270 g
Working environment
Temperature: 5℃~40℃
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 3.7V (rechargeable lithium batteries)
Input power: ≤1.5VA

Write Your Own Review
You're reviewing:CA10S Handheld Veterinary Sidestream Capnograph VET CO2 Patient Monitor ETCO2